PROVA 700 Milli-Ohmmeter
|
|
|
- Best resolution 1μΩ. Accuracy 0.05%
- 5A (120 mΩ) max. test current.
- Manual or Auto Range. (There are 6 ranges of current measurement, and furthermore 3 ranges in each current range.)
- Measurement of resistive and inductive materials (Two terminal measurement; Four terminal Kelvin measurement)
- Programmable Hi-Lo alarm with memory of 20 data.
- Auto-judgments of Hi (readings higher than Hi-Limit), Lo (readings lower than Lo-Limit) and Pass (readings between Hi-Lo Limits). Alarm (beeper) for Hi and Lo.
- Readings Hold function.
- Cable length measurement. The length unit can be selected between Meter (M) and Feet (FT).
- Resistance measurement results excluding cable resistances.
- Memory of 3,000 measurement data.
- Large LCD (5-digit) with backlight (users can choose to disable it).
- Rechargeable batteries and AC adaptor.
- Low battery (voltage) alarm.
- Low power consumption.
- Communicate with PC via RS232C (to USB bridge) cable.
- Application: various low-resistance (cables) measurement; contact resistance measurement, …etc.
|
|
|
Electrical Specifications (23℃±5℃)
Manual Range:
| Range |
Resolution |
Accuracy |
|
5A
|
1.000 mΩ ~ 8.000 mΩ
|
1 uΩ |
±0.05%±20 uΩ |
| 5.00 mΩ ~ 32.00 mΩ |
10 uΩ
|
±0.05%±200 uΩ
|
| 10.00 mΩ ~ 120.00 mΩ |
10 uΩ
|
±0.05%±200 uΩ
|
| 1A |
4.00 mΩ ~ 40.00 mΩ |
10 uΩ |
±0.05%±200 uΩ |
| 15.00 mΩ ~ 160.00 mΩ |
10 uΩ |
±0.05%±200 uΩ |
| 50.00 mΩ ~ 600.00 mΩ |
10 uΩ |
±0.05%±200 uΩ |
| 100 mA |
0.0400 Ω ~ 0.4000 Ω |
100 uΩ |
±0.05%±2 mΩ |
| 0.1500 Ω ~ 1.6000 Ω |
100 uΩ |
±0.05%±2 mΩ |
| 0.5000 Ω ~ 6.0000 Ω |
100 uΩ |
±0.05%±2 mΩ |
| 10 mA |
0.400 Ω ~ 4.000 Ω |
1 mΩ |
±0.05%±20 mΩ |
| 1.500 Ω ~ 16.000 Ω |
1 mΩ |
±0.05%±20 mΩ |
| 5.000 Ω ~ 60.000 Ω |
1 mΩ |
±0.05%±20 mΩ |
| 1 mA |
4.00 Ω ~ 40.00 Ω |
10 mΩ |
±0.05%±200 mΩ |
| 15.00 Ω ~ 160.00 Ω |
10 mΩ |
±0.05%±200 mΩ |
| 50.00 Ω ~ 600.00 Ω |
10 mΩ |
±0.05%±200 mΩ |
| 100 uA |
0.0400 kΩ ~ 0.4000 kΩ |
100 mΩ |
±0.05%±2 Ω |
| 0.1500 kΩ ~ 1.6000 kΩ |
100 mΩ |
±0.05%±2 Ω |
| 0.5000 kΩ ~ 6.0000 kΩ |
100 mΩ |
±0.05%±2 Ω |
|
Auto Range:
| Range |
Resolution |
Accuracy |
|
5A
|
1.000 mΩ ~ 8.000 mΩ
|
1 uΩ
|
±0.05%±20 uΩ
|
|
8.00 mΩ ~ 120.00 mΩ
|
10 uΩ
|
±0.05%±200 uΩ
|
|
1A
|
4.00 mΩ ~ 600.00 mΩ
|
10 uΩ
|
±0.05%±200 uΩ
|
|
100 mA
|
0.0400 Ω ~ 6.0000 Ω
|
100 uΩ
|
±0.05%±2 mΩ
|
|
10 mA
|
0.400 Ω ~ 60.000 Ω
|
1 mΩ
|
±0.05%±20 mΩ
|
|
1 mA
|
4.00 Ω ~ 600.00 Ω
|
10 mΩ
|
±0.05%±200 mΩ
|
|
100 uA
|
0.0400 kΩ ~ 6.0000 kΩ
|
100 mΩ
|
±0.05%±2 mΩ
|
|
|
General Specifications:
|
Battery Type:
|
Rechargeable, 2500mAh (1.2V) x 8 |
|
Battery Recharging Time:
|
10 hours |
|
AD Adaptor:
|
12V DC 2A |
|
LCD Display:
|
4 5/6 Digit LCD + Backlight |
|
Dimension:
|
257 (L) x 155 (W) x 57 (H) mm |
|
Weight:
|
v1160g / 40.0oz (Batteries included) |
|
Operation Environment:
|
-10℃ ~ 50℃, 85% RH |
|
Storage Environment:
|
-20℃ ~ 60℃, 75% RH |
|
Accessories:
|
Users manual x 1
RS232C (to USB Bridge) Cable x 1
Rechargeable batteries x 8
AC adaptor x 1
Software CD x 1
Software Manual x 1
Kelvin Clips x 1 set |
|
|
|
|
|
|
|