Model
|
ETCR3000
|
|
Function
|
Earth resistance measurement, Earth voltage measurement, Low resistance measurement
|
|
Power Supply
|
DC 9V (Zi-Mn dry battery R14S 1.5V 6 PCS, continuous standby for 300 hours )
|
|
Measuring Mode
|
Precise 3-pole measurement, simple 2-pole measurement
|
|
Measuring Method
|
Earth ground resistance: constant current change-pole method, measurement current 3mA Max,820Hz;
Voltage to ground: average value rectification
|
|
Data Mode
|
Average value, maximum, minimum
|
|
Wire Resistance Verification
|
Avoid error caused by testing wire such as failed to be fully inserted into tester interface or by poor contact or by user’ s replacing or lengthening testing wires, making it more accurate for earth ground resistance measurement.
|
|
Display Mode
|
4-digital super-large LCD display, blue screen backlight, LCD dimension: 128mm×75mm;Display field:124mm×67mm
|
|
Measuring Indication
|
During measurement, LED flash indicator, LCD count down display
|
|
Earth Ground Resistance
|
Range:0.01Ω~2000Ω;Resolution:0.01Ω;Accuracy:±1%rdg±3dgt
|
|
Earth Ground Voltage
|
Range:0V~600V AC;Resolution:1V;Accuracy:±1%rdg±3dgt
|
|
Tester Dimension
|
LWH:215mm×190mm×95mm
|
|
Testing Wires
|
Three wires: each for Red 20m, Yellow 10m, and Green 5m
|
|
Simple Testing Wire
|
2 wires: each for Red 1.5m and Green 1.5m
|
|
Auxiliary Earthing Rod
|
2 PCS: Φ10mm×150mm
|
|
Measuring Rate
|
Earth voltage measurement: 3 times/second; Earth resistance: 5 seconds/time
|
|
Measuring Times
|
Over 5000 times (20Ω range measuring 10Ω for one time and repeat after 25 seconds)
|
|
Circuit Voltage
|
Measuring voltage to ground: measuring below AC 600V
|
|
RS232 Interface
|
Possess RS232 interface, software supervision, storage data can be uploaded to computer, preserved and printed.
|
|
Communication Wire
|
One piece of RS232 communication wire, with length 1.5m
|
|
Data Storage
|
400 sets, flash display “FULL”icon to indicate storage is full
|
|
Data Hold
|
Data hold function: “HOLD”icon display
|
|
Data Read
|
Data read function: “READ” icon display
|
|
Overflow Display
|
Exceeding measuring range overflow function: “OL”icon display
|
|
Alarm Function
|
When measuring value exceeds alarm setting value, there is “Toot-toot-toot” alarm hint
|
|
Battery Voltage
|
When battery voltage decreases to about 7.8V, battery voltage low icon “ ”will display, reminding to replace battery.
|
|
Power Consumption
|
Backlight:25mA Max;Standby: 25mA Max(Backlight shut off);Measurement:70mA Max(Backlight shut off)
|
|
Weight
|
1443g(with batteries)
|
|
Working Temperature & Humidity
|
-10℃~40℃;below 80%rh
|
|
Storage Temperature & Humidity
|
-20℃~60℃;below 70%rh
|
|
Overload Protection
|
Measuring earth ground resistance: between each interfaces of E-P、E-C, AC 280V/3 seconds
|
|
Insulation Resistance
|
Over 10MΩ(between circuit and enclosure it is 500V)
|
|
Withstanding Voltage
|
AC 3700V/rms (Between circuit and enclosure)
|
|
Electromagnetic Features
|
IEC61010-4-3,Wireless frequency electromagnetic field≤1V/m
|
|
Protection Type
|
IEC61010-1、IEC1010-2-31、IEC61557-1,5、IEC60529(IP54)、CAT Ⅲ 300V 、Pollution etc.
|
|
Accessories
|
Tester: 1PC; Earthing Rod: 2PCS; Testing Wire: 3PCS; Simple testing wire: 2PCS; RS232 Com-cable: 1PC; Software disk: 1PC; Tester bag: 1PC
|